3. DEFINITIONS
Post-Mortem Forensic Toxicology -determines the absence or
presence of drugs and their metabolites, chemicals such as ethanol
and other volatile substances, carbon monoxide and other gases,
metals, and other toxic chemicals in human fluids and tissues, and
evaluates their role as a determinant or contributory factor in
the cause and manner of death;
Human-Performance Forensic Toxicology - determines the absence
or presence of ethanol and other drugs and chemicals in blood,
breath or other appropriate specimen(s), and evaluates their role
in modifying human performance or behaviour. (The analysis of
ethanol in breath, although important, was not considered by the
committee because such tests are not conducted in a laboratory
setting); and
Forensic Urine Drug Testing - determines the absence or
presence of drugs and their metabolites in urine to demonstrate
prior use or abuse.
Standard -a reference material
possessing one or more properties that are sufficiently well
established that it can be used to prepare calibrators.
Calibrator - a solution , either
prepared from the reference material or purchased, used to
calibrate the assay. Where possible, calibrators should be
prepared in a matrix similar to that of the specimens.
Control - a solution either prepared
from the reference material (separately from the calibrators; that
is, weighed or measured separately), purchased, or obtained from a
pool of previously analyzed samples. Controls from any of these
sources are used to determine the validity of the calibration;
that is, the stability of a quantitative determination over time.
Where possible, controls should be matrix-matched to specimens and
calibrators, as indicated above.
Reference Material (RM)
-a material or substance one or more properties of which are
established sufficiently well to be used for calibration of an
apparatus, assessing a measurement or assigning values to
material. (AOAC Official Methods of Analysis (1984)).
Certified Reference Material (CRM) - a reference material,
one or more of whose properties are certified by a valid
procedure, or accompanied by or traceable to a certificate or
other documentation which is issued by a certifying body. (AOAC
Official Methods of Analysis (1984)).
4. PERSONNEL
4.1 Laboratory
Director
4.1.1 The forensic
toxicology laboratory should be directed by a person who is
qualified by reason of appropriate education and experience to
assume the required professional, organizational, educational,
managerial and administrative responsibilities.
4.1.2 That education
and experience should be comparable to those of persons certified
as Diplomates by the American Board of Forensic Toxicology.
4.1.3 Alternative
acceptable qualifications include a doctoral degree in one of the
natural sciences and at least three years of full-time laboratory
experience in forensic toxicology; or a Master's degree in one of
the natural sciences and at least five years of full-time
laboratory experience in forensic toxicology; or a Bachelor's
degree in one of the natural sciences and at least seven years of
full-time laboratory experience in forensic toxicology.
4.1.4 The director
should also have documented training and/or experience in the
forensic applications of analytical toxicology (such as court
testimony, research, participation in continuing education
programs, and/or peer review of appropriate manuscripts in the
field), including a knowledge of evidentiary procedures that apply
when toxicological specimens are acquired, processed, and stored
and when toxicological data are submitted as part of a legal
proceeding.
4.1.5 The laboratory
director should be responsible for ensuring that the laboratory
personnel are adequately trained and experienced to conduct the
work of the laboratory
4.1.6 The laboratory
director should be responsible for maintaining the competency of
laboratory personnel by monitoring their work performance and
verifying their skills. This training and experience should be
documented.
4.1.7 The laboratory
director should be responsible for the development of a complete,
up-to-date procedures manual that is available to and followed by
all personnel performing tests.
4.1.8 The laboratory
director should establish a procedure for validating new
analytical methodologies, and for maintaining a quality assurance
program to ensure the proper performance and reporting of all test
results.
4.1.9 Since forensic
toxicology laboratories handle controlled substances and generate
results essential to the criminal justice system, the director, to
the extent practical or permitted by law, should exert reasonable
efforts to ensure that all personnel meet high ethical and moral
standards.
4.2 Other Laboratory
Staff
The
range and type of duties of other laboratory personnel will vary
according to the size and the scope of the laboratory. It is
recommended that each laboratory should have the following.
4.2.1 A person with the
title of deputy director, assistant laboratory director, assistant
chief toxicologist, or supervisory toxicologist, who has
sufficient training and experience to be familiar with all
administrative and testing procedures. He or she may supervise
the work of all analysts, and should be capable of performing full
scientific review of all test data, and of acting for the
laboratory director in the director's absence. It is recommended
that such individuals should have a minimum of a Bachelors degree
in a natural science and 3 years of training in analytical
toxicology, at least 1 year of which is in forensic toxicology.
4.2.2 One or more
technicians who are capable of performing a variety of test
procedures for alcohol, drugs, and other chemicals. A technician
may supervise and review the work of less experienced technicians,
and may supervise a section in a larger laboratory. It is
recommended that such individuals should have a minimum of a
Bachelor's degree in a natural science, at least 1 year of
experience in analytical toxicology and 6 months experience in the
present employment.
4.2.3 One or more
analysts who are capable of performing tests for one or several
analytes, and who are certified in each procedure by the
laboratory director. These analysts may be limited in function to
perform specified tasks - for example, an analyst who performs
only immunoassays.
5. STANDARD OPERATING PROCEDURES
5.1 The laboratory
should have a standard operating procedure manual (SOP) that is
complete, up-to-date, and available to all personnel who are
performing tests.
5.2 The SOP manual
should include detailed descriptions of procedures for sample
receiving, accessioning, chain-of-custody, analysis, quality
assurance and quality control, review of data, and reporting.
5.3 The SOP manual
should include administrative procedures as well as analytical
methods and be reviewed, signed, and dated whenever it is first
placed into use or changed.
5.4 The SOP manual
should include, for each analytical procedure if appropriate, the
following: a) theory and principle of the method, b) instructions
for preparation of reagents, c) details of the analytical
procedure, d) instructions for preparation of calibrators and
controls, e) information about any special requirements for
handling reagents or for ensuring safety, f) validation parameters
(e.g. LOQ, linearity), g) criteria for the acceptance or rejection
of qualitative or quantitative results and h) references.
5.5 When the required
documentation is not available for infrequently performed assays,
it should be added as each is performed for the first time.
5.6 The laboratory should
maintain out-dated copies of the SOP manual and provide a means
for their retrieval from archival storage.
6. SAMPLES AND RECEIVING
6.1 Specimen Collection
and Labelling
The
proper selection, collection, and submission of specimens for
toxicological analyses is of paramount importance if analytical
results are to be accurate and their subsequent interpretation is
to be scientifically sound and therefore useful in the
adjudication of forensic cases. These guidelines can apply equally
to investigations by Medical Examiners or Coroners (postmortem
forensic toxicology) and to investigation by law-enforcement
agencies of drivers suspected of being under the influence of
alcohol or other drugs.
6.1.2 The director
should develop and provide detailed guidelines and instructions to
all agencies or parties the laboratory serves.
6.1.3 Instructions
should state the types and minimum amounts of specimens needed to
accomplish the requisite analyses and subsequent interpretations.
6.1.4 Whenever
possible, the amount of specimen collected should be sufficient to
ensure that enough remains for subsequent re-analysis if required.
6.1.5 Instructions
should include specific requirements for the type and size of
specimen containers and, if appropriate, the type and amount of
preservative to be added to biological fluids.
6.1.6 Instructions for
labelling individual specimen containers, and acceptable
conditions for packing and transportation, should also be
provided.
6.1.7 Submitting
agencies should be instructed on how to label clearly all samples
from living subjects or decedents who may carry a highly
infectious disease such as tuberculosis, hepatitis or Human
Immunodeficiency Virus.
6.1.8 Each specimen
should be identified as to type. For blood, the anatomical site
of collection should be stated. When antemortem and/or perimortem
specimens are available from a decedent, each specimen should be
labelled with the time and date of collection.
6.1.9 The name of the
subject from whom the specimens were collected should appear on
each label together with other appropriate identification; for
example, the Medical Examiner's Case Number and/or the subject=s
Social Security Number.
6.1.10 Where provided, the time and date registered
for each specimen should be initialled or signed by a responsible
person who performed or witnessed the collection and who assumes
responsibility for the chain of custody.
6.2 Specimen Handling
6.2.1 A
chain-of-custody form should be designed that will accompany
specimens from the place of collection to the laboratory. This
document may be incorporated in the laboratory-request form.
6.2.2 Handling and
transportation of a specimen from one individual or place to
another should always be properly documented.
6.2.3 The
chain-of-custody section should be properly completed by
responsible personnel at the time the specimens are collected.
6.2.4 Every effort
should be made to minimize the number of persons handling a
specimen.
6.2.5 Individual
specimens should be transported and stored in such a manner as to
minimize the possibility of degradation, contamination, tampering
and/or damage in shipment.
6.2.6 The condition of
the external package should be documented upon receipt at the
laboratory, either on the requisition form that accompanies the
specimen(s), in the log book, on the external chain-of custody
form, or on other documents that constitute normal laboratory
records.
6.2.7 Acceptable means
of transporting specimens to the laboratory may include
hand-delivery, national postal service, or a private or government
courier service.
6.3 Specimen Receipt
6.3.1 The means of
delivery of specimens should be recorded by the receiving
laboratory.
6.3.2 Shipping
containers should be opened only in a secure area and only by an
individual designated to record receipt of specimens. A "secure
area" may be defined as an area to which unauthorized individuals
do not have access without escort by authorized personnel.
6.3.3 A hard copy of
the specimen-receipt record should be permanently maintained; this
record may be computer-generated, typed, or hand-written.
6.3.4 Specimens should
be logged-in immediately upon receipt.
6.3.5 The integrity of
the individual specimen container should be checked as should the
condition of each specimen. Discrepancies should be recorded.
6.4 Recommended
Amounts of Specimens
6.4.1 Post-Mortem Forensic Toxicology Specimens: In death
investigations, the types and minimum amounts of tissue specimens
and fluids needed for toxicological evaluation of the role of
drugs and other toxic chemicals are frequently dictated by the
analyte or analytes that must be identified and quantitated. For
example, in the evaluation of carbon monoxide poisoning, 10 mL of
whole blood would be sufficient with adequate specimen volume
remaining for re-analysis if required. On the other hand, to
evaluate the role of amitriptyline in a death, 25 gm of liver, 10
mL of heart blood, 5 mL of peripheral blood and the entire gastric
contents should be made available.
Many
deaths involve ingestion of multiple drugs, necessitating larger
amounts of tissue and fluids to be collected at autopsy for
toxicological examination. The following is a suggested list of
specimens and amounts to be collected at autopsy in such cases:
Brain 100 gm
Liver 100 gm
Kidney 50 gm
Heart
Blood 25 mL
Peripheral Blood 10 mL
Vitreous Humor All Available
Bile All Available
Urine All Available
Gastric Contents All Available
Unique poisons may dictate the need for other specimens, e.g. lung
and intestine. Such cases should be addressed on an individual
basis. However, the amount of specimen routinely collected should
be sufficient to allow re-analysis for one or more analytes at a
later time, should the need arise.
6.4.2 Human Performance Forensic Toxicology Specimens: As defined
earlier, this activity encompasses the identification and
quantitation of ethanol and other drugs and chemicals in blood,
breath or other appropriate specimens for evaluation of their role
in modifying human performance and behaviour. The analysis of
breath ethanol was not considered by this Committee.
Although in many instances the analytes are
clearly specified in advance in human performance forensic
toxicology testing, the spectrum of drugs and chemicals may
potentially approach those encountered in postmortem toxicology.
Because of this and remembering the difficulties involved in
obtaining samples from living persons, it is recommended that a
minimum of 15 mL of blood be collected for toxicological analysis.
Urine may also be submitted for
testing; a minimum volume of 30 mL is recommended. It must be
emphasized that neither qualitative nor quantitative analysis of
urine permits an evaluation of the effect of the drug or chemical
on human behaviour. If other specimens are submitted and
analyzed, any conclusions regarding drug use or effects on human
behavior should be based only on appropriate validated scientific
studies.
7. SECURITY AND CHAIN-OF-CUSTODY
7.1 The Laboratory
7.1.1 Access to the
forensic toxicology laboratory should be limited. The laboratory
director should authorize and document the personnel able to enter
designated areas.
7.1.2 Unauthorized
personnel should be escorted and may be required to sign a
log-book upon entry and departure from the laboratory, recording
the time, date and purpose of the visit.
7.1.3 The physical
layout of the laboratory must be such that unauthorized personnel
cannot enter without detection.
7.2 Specimens
7.2.1 Receipt of
specimens should be indicated by signature, and date and time of
individuals receiving the specimens.
7.2.2 Specimens
received should be labeled with the name of the decedent or
suspect, case number, specimen type (e.g. blood) or unique
identifier, date specimen taken and identification of the
individual taking the sample.
7.2.3 Specimens must
be stored in a secure manner.
7.2.4 For the
maintenance of specimen security it is recommended that, where
possible, the laboratory have a separate accessioning area. In
this area, specimens are received, assigned accession numbers,
aliquots removed and/or stored in refrigerator/freezers.
7.2.5 Any transfer of
specimens, or portions thereof that are removed for analysis, must
be documented as part of the permanent laboratory record.
7.2.6 It is
recommended that chain of custody documentation reflects not only
the receipt of the specimen from an outside source, but also
transfers of the specimen or an aliquot thereof, within the
laboratory. If multiple specimens are involved, a batch form may
be used.
7.2.7 An aliquot or a
batch of aliquots chain of custody may be used for indicating the
transfer of portions of specimens for testing. This form should
indicate the date, the test for which the aliquot was taken, the
laboratory accession numbers, the identity of the individual
obtaining the aliquots and the identity of the individual to whom
the aliquots were given, if applicable.
7.2.8 Specimens may be
transferred to a secure long-term refrigerator/freezer after
analysis. Transfers between storage areas and/or subsequent
disposal should be documented. The laboratory should develop a
standing operating procedure for retention and disposal of
specimens. This procedure should reflect local, state, or federal
regulations.
7.2.9 The laboratory
should maintain a written policy and instructions pertaining to
retention, release and disposal of specimens.
8. ANALYTICAL PROCEDURES
8.1 Confirmatory
Tests
8.1.1 As a general
matter of scientific and forensic principle, the detection or
initial identification of drugs and other toxins should be
confirmed whenever possible by a second technique based on a
different chemical principle.
8.1.2 Where possible,
the confirmatory (second) test should be more specific and
sensitive than the first test for the target analyte. The use of
mass spectrometry is recommended as the confirmatory technique,
where practical.
8.1.3 In some
circumstances, confirmation using the same GC system as the first
might be acceptable if chemical derivatization (e.g. silylation or
acylation) is used to change the retention times. However,
confirmation using a second GC system with a similar though not
identical column, is not usually acceptable since the retention
indices of many analytes may not differ substantially from one
system to the other (e.g. DB-1 and DB-17).
8.1.4 Use of a second
immunoassay system (e.g. RIA) to confirm another immunoassay
(e.g. FPIA) is not regarded as acceptable, even though the assays
differ somewhat in principle. The rationale for this is that the
analytes which cross-react with one assay are also likely to
cross-react in the second assay because the antibodies may be
raised to the same drug or closely related substance.
8.1.5 A second
immunoassay with different cross-reactivity may sometimes be used
to augment the initial screen (for example a broadly
cross-reacting opiate immunoassay, followed by a second
immunoassay with more specific cross-reactivity to unconjugated
morphine). These results would normally still require
confirmation with a more specific method (e.g. GC/MS).
8.1.6 It is a good
practise to confirm the identity of an analyte in a different
extract of the same specimen from that used for the first test, or
in a second specimen. However, confirmation of a drug or toxin in
the same original extract of a single specimen would not normally
be regarded as acceptable, since that would not rule out the
possibility that the extract became contaminated during the
extraction.
8.1.7 The quantitation
of an analyte may serve as acceptable confirmation of its identity
if it was initially detected by a significantly different method
(e.g. GC/MS SIM quantitation of a drug detected by immunoassay).
8.1.8 Where mass
spectrometry is used in selected ion monitoring mode for the
identification of an analyte, whether as part of a quantitative
procedure or not, the use of at least one qualifying ion for each
analyte and internal standard, in addition to a primary ion for
each, is strongly encouraged where possible. Commonly used
acceptance criteria for ion ratios is
"20%
relative to that of the corresponding control or calibrator.
However, it is recognized that some ion ratios are concentration
dependent and that comparison to a calibrator or control of
similar concentration may be necessary, rather than an average for
the entire calibration. Ion ratios for LC/MS assays may be more
concentration and time dependent than for GC/MS and therefore
acceptable ion ratio ranges of up to
"25%
or 30% may be appropriate.
8.1.9 GC/MS chemical
ionization and LC/MS mass spectra are often simpler than GC/MS EI
spectra and therefore afford less options for the choice of
qualifier ions. However, it is often possible to adjust the
ionization energy (e.g. cone or fragmentor voltage in LC/MS) in
order to produce additional or stronger secondary ions. Running
the sample under conditions of both weak ionization (to maximize
the quantitation ion signal) and stronger ionization (to promote
fragmentation and facilitate confirmation of identity) is an
option.
8.1.10 It is
recommended that at least the presence of a drug or toxin be
verified in more than one specimen, or if only one specimen is
available by replicate analyses on different occasions and with
adequate positive and negative controls in the same matrix.
However, it is acknowledged that an analyte will not necessarily
be present in all specimen types.
8.1.11 Use of a second
confirmatory technique is encouraged for all analytes, including
ethanol
(e.g. GC, ADH, or colorimetric) and carbon
monoxide (e.g. visible spectrophotometry, palladium chloride or
GC).
8.1.12 It is recognized
that in some circumstances a suitable second test procedure is not
available and the probability that the first test is incorrect is
almost zero. For example, the probability that a 75%
carboxyhemoglobin in a well-documented suicide is incorrect, when
obtained by a properly conducted spectrophotometric assay, is
exceedingly low. However, the unexpected finding of a 30%
carboxyhemoglobin from a motor vehicle accident victim by a
similar determination in blood holds a lower degree of certainty.
8.2 Method
Calibration and Validation
8.2.1 When conducting
analyses, laboratories may group specimens into batches. Each
batch should contain a sufficient number of calibrators and
controls, the total number of which will depend on the size of the
batch and the nature of the tests.
8.2.2 When analyses are
being performed on unusual specimens (decomposed tissue, vitreous
fluid, etc.), appropriate matrix-matched calibrators should, when
possible, be prepared and tested concurrently with the specimens.
8.2.3 For immunoassays,
a laboratory should, at a minimum, be able to demonstrate that the
blank or negative calibrator plus two standard deviations does not
overlap with the cut-off or the lowest positive calibrator.
Alternatively, the laboratory may determine the limit of detection
(LOD) by determining the mean value for the blank and adding three
standard deviations to this value (LOD = Xm + 3SD). However, it
should be noted that for other assays (e.g. GC, HPLC) the true LOD
may be higher than indicated by this formula if significant
adsorption or other losses occur. For example, in chromatographic
assays, the LOQ might be the smallest blood concentration of a
drug needed to give a peak height three times the noise level of
the background signal from a blank blood sample. Alternatively,
for infrequently performed assays where the analyte measured is
always within the calibration range of the assay and well above
the LOD, it may be sufficient to indicate that the detection limit
is Aless than@a certain value. Thus the true LOD may be derived
experimentally, but should not be less than the blank plus three
standard deviations. The limit of quantitation (LOQ) may be
derived by adding ten standard deviations to the true value of the
blank. However, it is preferable to determine the LOQ
experimentally as the lowest concentration for which an acceptable
coefficient of variation can be routinely achieved.
8.2.4 For
chromatographic assays, the LOD and LOQ may be administratively
defined in terms of the concentration of the lowest calibrator,
and therefore may not need to be determined experimentally.
However, if results are reported below the value of the lowest
calibrator, LOD and LOQ should be determined.
8.2.5 The use of a
suitable internal standard for all chromatographic assays (e.g.
GC, HPLC, GC/MS) is recommended. The internal standard should
have chemical and physical properties as similar to the analyte as
possible. If the analyte is to be derivatized, an internal
standard should be chosen which will form an analogous
derivative. Stable isotope (e.g. deuterated) standards are
recommended for GC/MS assays, although well chosen non-deuterated
internal standards may occasionally give equivalent or better
performance. The internal standard should be added to the sample
at the earliest possible stage in the method, and in any event
before buffering and extraction of the sample. Markers which are
added after the initial extraction are regarded as "external
standards" and are discouraged.
8.2.6 Linearity of the
procedure should be established by typically using at least three
calibrators. The concentration of the calibrators should be such
that they bracket the anticipated concentration of the specimen(s).
If the concentration of the specimen exceeds the concentration of
the highest calibrator, the specimen should be diluted and
re-extracted if accurate quantitation is required. Otherwise the
specimen should be reported as having a concentration greater than
the highest calibrator. If the concentration of the specimen
should be less than that of the lowest calibrator, an additional
calibrator should be set up which falls below the expected range
of the analyte in the sample. Alternatively, the volume of the
specimen may be doubled and re-extracted if it can be demonstrated
that the assay is not matrix dependent. If an accurate
quantitation is not necessary, then the specimen can be reported
as containing the analyte at less than the lowest calibrator (as
an alternative to the term "trace amount"). It is acknowledged
that some assays are inherently non-linear and that the use of
quadratic or other mathematical models may be necessary.
8.2.7 Criteria for
acceptance of a chromatographic calibration should be stated in
the method. For a multi-point calibration this factor is usually
the correlation coefficient. For most applications, an acceptable
correlation coefficient is 0.99. However, there may be
circumstances where a correlation coefficient of 0.98 is minimally
acceptable. In addition, it is good practice to evaluate the
range of the calibration by calculating the value of each
calibrator against the curve.
Values of "20%
are generally acceptable for most applications, although
"10% are preferred for
analytes such as ethanol. Single point calibrations are
discouraged unless controls are used at or close to the upper and
lower quantitative reporting limits.
8.2.8 For specimens
having concentrations significantly higher than the highest
calibrator, the laboratory should exercise precautions so that
carry-over of analyte into the next specimen does not occur.
Similarly, specimens with very low concentrations should be
checked to ensure that carry-over from a previous very high
positive has not occurred.
8.2.9 It is recognized
that for a variety of reasons occasional analytical results will
be Aoutliers@; that is, analytical values which deviate
significantly and spuriously from the true value. "Outlier"
results of control, blanks or calibrators should be obvious.
However Aoutlier@results of case specimens may not be identified
if only run singly, unless that result can be compared with one
from a separate analytical determination. For this reason
replicate extraction and quantitative analysis, at least in
duplicate, is recommended. The laboratory should determine the
acceptable criteria for replicate analysis. A maximum deviation
of
"20%
of the mean is recommended.
8.2.10 Retention time
should be part of the acceptance criteria for chromatographic
assays. For GC based assays, deviations of 1 - 2% from the
calibrators or controls may be acceptable. Slightly larger
deviations may be acceptable for HPLC based assays, particularly
where the mobile phase is being programmed non-isocratically.
8.3 Method of
Standard Additions
It
is recognized that the matrix of some forensic specimens may be
"unique" in some way (e.g. putrefied or embalmed) such that it is
difficult or impossible to obtain a similar matrix for the
preparation of reliable calibrators and controls. In these
circumstances, the use of a "standard addition" procedure may be
preferable to a conventionally calibrated assay. In the method of
"standard addition" known amounts of analytes are added to
specimen aliquots and quantitation performed by comparing the
proportional response of the fortified aliquots with that of the
unknown specimen. Use of an internal standard and a multiple
point calibration is strongly recommended.
9. QUALITY ASSURANCE AND QUALITY
CONTROL
9.1 Quality
Assurance
9.1.1 Quality
assurance encompasses all aspects of the analytical process, from
specimen collection and reception through analysis, data review
and reporting of results. It includes, but should not be limited
to, quality control of each analysis and proficiency testing of
the laboratory.
9.1.2 Quality
assurance assumes a unique role in the forensic science
disciplines because results are subject to challenge in the
Aadversarial@ justice system. One purpose of a quality assurance
program is to detect error, whether random or systematic, and to
initiate appropriate remedial action.
9.1.3 Standards used
should be appropriate for the test being performed, and
documentation should be maintained describing their sources and
dates of acquisition. Reference material should be stored so as
to ensure its stability and integrity. If a standard is prepared
in the laboratory, the source(s) of the chemical reagent(s), the
method of preparation, and verification of the final product
should be recorded and maintained on file.
9.1.4 Where practical,
the identity and purity of reference materials should be verified
by the laboratory.
9.1.5 Labelling should
be uniform for all standards and reagents. Date of acquisition or
preparation, and the initials of the preparer, should be included
on the label. The expiration date should always appear on the
label. An expiration date furnished by a vendor/manufacturer
determines the useful lifetime of the standard/control unless it
can be verified beyond that date.
9.1.6 Initially, a
sufficient number of calibrators should be run to determine the
characteristics of the calibration curve; a blank and at least
three calibration points are recommended for the initial
calibration process. The stability of the calibration curve should
be tested under laboratory conditions by the addition of controls,
both positive and negative.
9.1.7 Controls are not
analyzed for calibration purposes. As a general rule an adequate
set of controls should include, at a minimum, a specimen that does
not contain the analyte (defined as a negative control) and a
specimen containing the analyte at a concentration that
realistically monitors the performance of the assay. Additional
controls can be used to test the linearity of the calibration over
the desired range.
9.1.8 The SOP manual
should specify corrective action to be taken when control results
are outside acceptable limits. Under optimal conditions a
laboratory should have a quality control supervisor, but having a
staff member dedicated to quality control may be impractical for
small laboratories.
9.1.9 Forensic
toxicology laboratories should participate in an external
proficiency testing program which includes, at a minimum, a
proficiency testing program for alcohol in blood or serum, and for
drugs in at least one type of specimen; the program should
realistically monitor the laboratory's quantitative capability.
9.1.10 The laboratory
director should regularly review results of quality control and
proficiency testing. Signing and dating the record constitutes
appropriate evidence of review. It is important that bench
personnel be informed of quality control and proficiency test
results. Attention should be given to procedures for monitoring
potential sources of error. Proficiency test materials should be
retained until the summary report is received and any corrective
action satisfactorily completed.
9.1.11 Appropriate and
timely corrective action in the event of proficiency test errors
is essential. False positive errors are the most serious and
possible causes of the error must be thoroughly investigated,
including contamination of glassware and carry-over. A false
negative result can be defined as failure to detect a substance
which the laboratory claims to be able to detect, or that should
have been detected by the method. By this definition, a false
negative indicates a failure that should be investigated
expeditiously. A false negative can also occur because the routine
methods of the laboratory will not detect the analyte at all, or
at the spiked concentration. In this instance the laboratory
director should decide whether the analytical procedures need
revising, or whether the failure to detect that analyte at the
spiked concentration is acceptable (e.g. the concentration is
below that of toxicological interest). All corrective action
should be documented.
9.1.12 Quantitative
proficiency test errors should also be investigated. Usually, the
target concentrations of analytes are expressed in terms of the
mean value for all participants in the survey, plus or minus 1 SD
or 2 SD. Occasionally, the weighed-in target may be disclosed.
Where the magnitude of an error is large, the need for corrective
action is obvious and the underlying cause may be easy to
determine. For some analytes, especially those infrequently
quantitated, 2 SD, a common measure of acceptability, may
represent an unacceptably large percentage deviation from the
mean. Therefore, a realistic percentage deviation should be used,
such as
"20%
or "30%.
Depending on the magnitude of the error, corrective action may be
as simple as review of the assay results to ensure that the
calibration was valid, that the assay was in control, and that any
transcriptions were accurate. For more serious errors, corrective
action may require repeating the analysis, re-validation of the
assay, or even redevelopment of the test. All corrective action
should be documented.
9.1.13 Routine
maintenance of equipment is an important part of any quality
assurance program.
It
is a good practice to document all routine and non-routine
maintenance, including tasks such as changing septa and liners on
GCs. Documentation may be in a logbook, which can be kept by
larger equipment, or check-sheets filed in a ring binder.
Multiple items of similar equipment
(e.g. pipettors) should be labelled in order to readily
differentiate them.
9.2
Quality Control
9.2.1 Control Materials: In the true sense, a control is a
test sample, identical to the unknown, but containing the analyte
at a known concentration. With each batch of specimens, whether a
single specimen or multiple ones, controls would be carried
through the procedure in parallel with the unknowns. It is
suggested that each batch of specimens include at least 10%
controls. The controls must include one positive and one
negative control. For qualitative assays positive and negative
controls, acceptable results may simply be Apositive@or Anegative@,
respectively. For quantitative assays, negative controls should
give results that indicate the analyte is absent, or below the LOD
for the assay. An acceptable positive control result of
"20% is recommended for most
drugs, except for controls that are at or close to the LOQ of the
assay, when "25-30% may be
more realistic. The control must give a result within a
predetermined deviation from its mean value, or the test is deemed
"out of control" and therefore, the result generated from the
unknown specimen is unacceptable.
9.2.2 It is a common
and accepted practice in clinical laboratory work to obtain or
prepare material and then establish the target range by replicate
analysis of the control in parallel with existing QC material.
For example, control material may be prepared by pooling specimens
from multiple cases. While that approach is still accepted in
forensic toxicology, it is scientifically less desirable than
preparing or purchasing control material with a specific
weighed-in target concentration, which will allow independent
verification of calibration accuracy. If control target ranges are
experimentally determined, it is important for that range to be
verified against control material, prepared commercially or
independently in-house, prior to it being put into routine use.
9.2.3 For some
forensic toxicology procedures, providing a true control is no
more difficult than any other test. For others, however, in which
the matrix may be unique (e.g. decomposed tissues, bone, hair or
nails) , providing a control is not only difficult, but can never
approach the ideal of being identical to the unknown specimen.
Controls should be prepared from standard material from a
different source than that used in calibration of the assay.
Where this is not practical, the control should at least be
prepared using a different weighing or dilution than that used to
prepare the calibrators. Control material prepared from the same
solution used to prepare the calibrators is unacceptable, since
any errors made in preparation of the standard solution will not
be detected.
9.2.4 Open Controls:
Open controls are those whose identity and expected result are
known to the analyst. They can be purchased from commercial
vendors, prepared in the laboratory, distributed by professional
organizations or saved and pooled from former cases. Regardless
of the source, the concentration of the analyte in the control
must be validated.
For tissue specimens or other unusual matrices,
more innovative approaches may be necessary. Fortifying drug-free
matrices, such as tissue homogenates, out-dated blood bank blood,
plasma to simulate the unknown specimen is acceptable. A "blank"
or negative control may, of course, be the unfortified matrix.
9.2.5 Results from
quantitative quality control material should be recorded in a
manner that readily permits the detection of trends such as the
deterioration of reagents, calibrators or controls. For frequently
run controls, results may be plotted in a graphical manner such as
a Levy-Jennings plot. For less frequently run material, tabulation
is acceptable. Determination of the coefficient of variation for
the controls may give useful information about the precision of
the assay, and may indicate which assays need further development.
9.2.6 Blind
Controls: As the name implies, these are identical to open
controls except their identity is unknown to the analyst. It is
generally recognized that this is the ideal way to maintain
quality control. A blind control should test the entire
laboratory process including receiving, accessioning, analysis and
reporting. This can be accomplished by setting up a "dummy
account" or by co-operation with the submitting agency. Such
blind controls are sometimes called "double blinds". A more
practical approach is to have the accessioning section insert
blind controls into each batch of specimens. However, either of
these processes can be difficult to accomplish in a small
laboratory; they are both costly and time consuming.
9.2.7 Sources of
Information: There are many good references relating to the
general topic of quality control as well as to specific
application relating to analytical procedures. These include:
Tietz Textbook of Clinical Chemistry, 3rd edition,
by Carl A. Burtis, Edward R. Ashwood, Norbert W. Tietz, Adrianne
Williams (Editors), W.B. Saunders Co., Philadelphia, 1998;
Cost-Effective Quality Control: Managing the Quality and
Productivity of Analytical Processes, by James O. Westgard and
Patricia L. Barry, AACC Press, Washington DC, 1986; Quality
Assurance in Postmortem Toxicology by Wilmo Andollo, Chapter
12.5 in Handbook on Drug Abuse, Steven Karch senior editor,
W. Lee Hearn section editor, CRC, Boca Raton, FL, 1998, pp
953-969.
9.3 Reference
Materials
9.3.1 The National
Institute of Standards and Technology (NIST), formerly the
National Bureau of Standards (NBS) , refers to these as Standard
Reference Material (SRM). For example, a specific RM may have a
melting point of such sharpness and reproducibility that it can be
offered as an RM for the calibration of a thermometer in a melting
point apparatus. However, it may not be appropriate for preparing
a calibration curve. A CRM, or SRM, suitable for the preparation
of a standard to which calibration material can be compared, must
be certified by a method generally recognized by the scientific
community as one which validates the CRM for this purpose. The
nature of the procedure depends, of course, on the properties of
the analyte.
9.3.2 Several
different organizations supply CRMs of a variety of types. A
complete listing of organizations supplying CRMs is available from
the International organization for Standardization, American
National Standards Institute, 1430 Broadway, New York, NY 10018.
Catalogs or literature describing CRMs are
available from individual organizations, such as the NIST.
9.3.3 It is important
to remember that most RMs are not 100% pure. The label or package
insert should indicate the purity or the nature of the
contaminants or the degree of water of hydration. Further
instructions may provide guidance as to how the RM is to be used.
For example, perhaps it must be protected from light, or stored at
a low temperature or protected from moisture. These instructions
must be carefully followed in order to use the RM according to its
specifications.
9.3.4 Drugs, perhaps
more than many other chemicals, may have limited shelf-lives.
Degradation due to photo-reactions, oxidation in the air or by
other means, requires that periodic
assessment of these changes must be monitored.
Methods for detecting such changes are varied but even RMs may not
retain their original purity. RMs supplied in solution may have
more limited stability than those supplied as pure, dry, solids.
9.3.5 The importance of acquiring pure
chemicals used as standards and periodically monitoring their
purity, requires the development and implementation of procedures
which are part of the standard operating procedure of the
laboratory. The steps which can be used are summarized as follows:
1) maintain
instruments and all measuring devices at optimal performance with
regular calibration checks. 2) acquire chemicals to be used as
standards from reliable sources who validate the stated purity,
preferably by a certifiable trace to a CRM or SRM, or 3) acquire
chemicals as RM, carefully following any instructions accompanying
the RM for maintaining anhydrous conditions or to avoid
deterioration, or 4) acquire chemicals from other sources but
always assess the purity of the material by appropriate
measurement of physical constants and/or instrumental methods.
5)
regardless of the source of the chemical for preparation
of the standard, devise a means by which the standard can be
monitored periodically in order to detect any deviation from its
original purity.
6)
before using a newly prepared standard, compare its
properties with a previously validated standard or with a CRM or
SRM.
9.4 Metabolites
9.4.1 Many testing
procedures, particularly immunoassay tests, are targeted to detect
drug metabolites. As might be expected, these are more difficult
to obtain in pure form, free of interferences and certified as to
their authenticity. A number of commercial sources offer drugs and
some metabolites, together with deuterated forms useful as
internal standards in GC-MS. Frequently the commercial sources
will supply a statement of purity with the material. This is not
the same as a CRM or SRM, but after verification of purity, may be
quite acceptable.
9.4.2 Metabolites of
pharmaceutical drugs can, at times, be obtained from the company
which manufactures them. This often requires a personal contact
with an appropriate official of the company, completion of
necessary paperwork and some time delay. The Physicians' Desk
Reference in its "Manufacturers' Index" lists names and
telephone numbers of contact officials.
9.4.3 When the identity
of the metabolite has been described in a reputable scientific
journal, but no source is evident, a search of catalogs from
suppliers of organic chemicals may be fruitful. If this is not
successful, then it may be necessary to synthesize the
metabolite. In this case its identity should be confirmed by
standard, acceptable methods. In all of these alternatives,
purity must be assessed.
10. REVIEW OF DATA
10.1
Before results are reported, each batch of analytical data should
be reviewed by scientific personnel who are experienced with the
analytical protocols used in the laboratory. At a minimum this
review should include:
.*
chain-of-custody documentation
.* validity
of analytical data (e.g., shape and signal-to-noise ratio of
chromatographic peak) and calculations
.* quality
control data.
10.2 Where possible,
the results should be reviewed in the context of the case history,
autopsy findings and any relevant clinical data. The review should
be documented within the analytical record.
11. REPORTING OF RESULTS
11.1 General
Recommendations
11.1.1 Many, if not
most, forensic toxicology laboratories are an integral part of
state or local government supported, medico-legal investigative
agencies, or are associated with them. Each laboratory must
follow the mandates of the particular agency and/or governmental
sub-division when reporting results. Thus, while it is neither
possible nor desirable to suggest a uniform format for reports,
they should include all information necessary to identify the case
and its source, and should bear test results and the signature of
the individual responsible for its contents.
11.1.2 The following
recommendations are made:
1) name and/or identification number
2) laboratory identification number
3) name of submitting agency or individual
4) submitting agency=s identification number
5) date submitted
6) date of report
7) specimens tested
8) test results
9) signature of approving individual
11.1.3 Although most forensic toxicology reports are confidential
and often sensitive in content, some jurisdictions may treat the
report as an official public document. If the results are
confidential, every precaution should be exercised to ensure that
a properly authorized person receives the information when it is
transmitted by telephone, computer, FAX, or any other method
different from conventional delivery of a written report. Each
laboratory should formulate its own policy for retention or
release of information and for response to requests for its
documentation.
11.2 Terminology in Reports